Shape Optimization of an Open Photoacoustic Resonator

Said Ali Said El-Busaidy, Bernd Baumann, Marcus Wolff*, Lars Duggen

*Corresponding author for this work

Research output: Contribution to journalJournal articleResearchpeer-review

50 Downloads (Pure)

Abstract

Photoacoustic (PA) measurements with open resonators usually provide poor detection sensitivity due to signal leakage at the resonator opening. We have recently demonstrated three different approaches for modelling the photoacoustic signal of open resonators. In this work, one of the approaches is applied for the optimization of the geometry of the T-shaped resonator for improved signal strength and thus sensitivity. The results from the numerical optimization show an increase in the photoacoustic signal by a factor of approximately 7.23. They are confirmed using numerical methods other than the one applied for the optimization and by experimental measurement. The measurement shows an increase in the photoacoustic signal by a factor of approximately 2.34.
Original languageEnglish
Article number2571
JournalApplied Sciences
Volume11
Issue number6
Number of pages11
ISSN2076-3417
DOIs
Publication statusPublished - 2021

Fingerprint

Dive into the research topics of 'Shape Optimization of an Open Photoacoustic Resonator'. Together they form a unique fingerprint.

Cite this