Quality of cardiopulmonary resuscitation in out-of-hospital cardiac arrest before and after introduction of a mechanical chest compression device, LUCAS-2; a prospective, observational study

  • Tinne Tranberg*
  • , Jens F. Lassen
  • , Anne K. Kaltoft
  • , Troels M. Hansen
  • , Carsten Stengaard
  • , Lars Knudsen
  • , Sven Trautner
  • , Christian J. Terkelsen
  • *Corresponding author for this work
  • Aarhus University Hospital
  • Region Midtjylland
  • Falck AS

Research output: Contribution to journalJournal articleResearchpeer-review

Abstract

BACKGROUND: Mechanical chest compressions have been proposed to provide high-quality cardiopulmonary resuscitation (CPR), but despite the growing use of mechanical chest compression devices, only few studies have addressed their impact on CPR quality. This study aims to evaluate mechanical chest compressions provided by LUCAS-2 (Lund University Cardiac Assist System) compared with manual chest compression in a cohort of out-of-hospital cardiac arrest (OHCA) cases.

METHODS: In this prospective study conducted in the Central Denmark Region, Denmark, the emergency medical service attempted resuscitation and reported data on 696 non-traumatic OHCA patients between April 2011 and February 2013. Of these, 155 were treated with LUCAS CPR after an episode with manual CPR. The CPR quality was evaluated using transthoracic impedance measurements collected from the LIFEPAK 12 defibrillator, and the effect was assessed in terms of chest compression rate, no-flow time and no-flow fraction; the fraction of time during resuscitation in which the patient is without spontaneous circulation receiving no chest compression.

RESULTS: The median total episode duration was 21 minutes, and the episode with LUCAS CPR was significantly longer than the manual CPR episode, 13 minutes vs. 5 minutes, p < 0.001. The no-flow fraction was significantly lower during LUCAS CPR (16%) than during manual CPR (35%); difference 19% (95% CI: 16% to 21%; p < 0.001). No differences were found in pre- and post-shock no-flow time throughout manual CPR and LUCAS CPR. Contrary to the manual CPR, the average compression rate during LUCAS CPR was in conformity with the current Guidelines for Resuscitation, 102/minute vs. 124/minute, p < 0.001.

CONCLUSION: Mechanical chest compressions provided by the LUCAS device improve CPR quality by significantly reducing the NFF and by improving the quality of chest compression compared with manual CPR during OHCA resuscitation. However, data on end-tidal Co2 and chest compression depth surrogate parameters of CPR quality could not be reported.

Original languageEnglish
Article number37
JournalScandinavian Journal of Trauma, Resuscitation and Emergency Medicine
Volume23
ISSN1757-7241
DOIs
Publication statusPublished - 22. Apr 2015
Externally publishedYes

Bibliographical note

Publisher Copyright:
© Tranberg et al.; licensee BioMed Central.

Keywords

  • CPR quality
  • Mechanical chest compression
  • Out-of-hospital cardiac arrest (OHCA)
  • Resuscitation
  • Prospective Studies
  • Humans
  • Male
  • Out-of-Hospital Cardiac Arrest/therapy
  • Treatment Outcome
  • Time Factors
  • Heart Massage/instrumentation
  • Denmark
  • Signal Processing, Computer-Assisted
  • Female
  • Aged
  • Software

Fingerprint

Dive into the research topics of 'Quality of cardiopulmonary resuscitation in out-of-hospital cardiac arrest before and after introduction of a mechanical chest compression device, LUCAS-2; a prospective, observational study'. Together they form a unique fingerprint.

Cite this