Image quality in coronary computed tomography angiography: influence of adaptive statistical iterative reconstruction at various radiation dose levels

Research output: Contribution to journalJournal articleResearchpeer-review

Abstract

Background Computed tomography (CT) technology is rapidly evolving and software solution developed to optimize image quality and/or lower radiation dose. Purpose To investigate the influence of adaptive statistical iterative reconstruction (ASIR) at different radiation doses in coronary CT angiography (CCTA) in detailed image quality. Material and Methods A total of 160 CCTA were reconstructed as follows: 55 scans with filtered back projection (FBP) (650 mA), 51 scans (455 mA) with 30% ASIR (ASIR30), and 54 scans (295 mA) with 60% ASIR (ASIR60). For each reconstruction, subjective image quality was assessed by five independent certified cardiologists using a visual grading analysis (VGA) with five predefined image quality criteria consisting of a 5-point scale. Objective measures were contrast, noise, and contrast-to-noise ratio (CNR). Results The CTDIvol resulted in 10.3 mGy, 7.4 mGy, and 4.6 mGy for FBP, ASIR30, and ASIR60, respectively. Homogeneity of the left ventricular lumen was the sole aspect in which reconstruction algorithms differed with a decreasing effect for ASIR60 compared to FBP (estimated odds ratio [OR] = 0.49 [95% confidence interval (CI) = 0.32-0.76; P = 0.001]). Decreased sharpness and spatial- and low-contrast resolutions were observed when using ASIR instead of FBP, but differences were not statistically significant. Concerning objective measurements, noise increased significantly for ASIR30 (OR = 1.08; 95% CI = 1.02-1.14; P = 0.006) and ASIR60 (OR = 1.06; 95% CI = 1.01-1.12; P = 0.034) compared to FBP. Conclusion ASIR significantly decreased the subjectively assessed homogeneity of the left ventricular lumen and increased the objectively measured noise compared to FBP. Considering these results, ASIR at a reduced radiation dose should be implemented with caution.

Original languageEnglish
JournalActa Radiologica
Volume59
Issue number10
Pages (from-to)1194-1202
ISSN0284-1851
DOIs
Publication statusPublished - 2018

Keywords

  • Journal Article
  • radiation safety
  • image quality
  • Cardiac
  • technical aspects
  • CT angiography
  • iterative reconstruction
  • Data Interpretation, Statistical
  • Computed Tomography Angiography/methods
  • Humans
  • Middle Aged
  • Male
  • Cardiac-Gated Imaging Techniques
  • Radiation Dosage
  • Contrast Media
  • Triiodobenzoic Acids
  • Female
  • Coronary Angiography/methods
  • Radiographic Image Interpretation, Computer-Assisted/methods

Fingerprint

Dive into the research topics of 'Image quality in coronary computed tomography angiography: influence of adaptive statistical iterative reconstruction at various radiation dose levels'. Together they form a unique fingerprint.

Cite this