Abstract
We evaluated an infrared tympanic thermometer (Genius 3000A) by comparing it with parallel measurements with an electronic rectal thermometer (Philips HP 5316) on 121 patients admitted to a geriatric department. Rectal temperature was on average 0.14 degree C +/- (ISD) above the ear temperature. 95% of the differences are within the interval from -1.18 degrees C to 1.46 degrees C. The coefficient of determination was only 0.30. The tympanic thermometer, Genius 3000A, cannot be recommended for daily use on a geriatric ward.
Udgivelsesdato: 1998-Aug-31
Udgivelsesdato: 1998-Aug-31
| Translated title of the contribution | Evaluation of ear temperature measurements in a geriatric department |
|---|---|
| Original language | Danish |
| Journal | Ugeskrift for Læger |
| Volume | 160 |
| Issue number | 36 |
| Pages (from-to) | 5175-7 |
| Number of pages | 2 |
| ISSN | 0041-5782 |
| Publication status | Published - 1998 |
Keywords
- Aged
- Body Temperature
- Cross-Sectional Studies
- Ear
- Evaluation Studies as Topic
- Female
- Geriatric Nursing
- Humans
- Male
- Middle Aged
- Rectum
- Reproducibility of Results
- Thermography
- Thermometers
- Tympanic Membrane
Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver