T1 - Effect of optimal lesion preparation prior to implantation of a magnesium-based bioresorbable scaffold
T2 - 12-month results of the OPTIMIS study
AU - Hansen, Kirstine Nørregaard
AU - Maehara, Akiko
AU - Trøan, Jens
AU - Noori, Manijeh
AU - Hougaard, Mikkel
AU - Ellert, Julia
AU - Veien, Karsten Tange
AU - Junker, Anders
AU - Hansen, Henrik Steen
AU - Lassen, Jens Flensted
AU - Jensen, Lisette Okkels
PY - 2025/8
Y1 - 2025/8
N2 - Background: Lumen reduction after bioresorbable scaffold implantation has been reported. This study aimed to assess the influence of pre-dilatation with a scoring balloon versus a standard non-compliant balloon prior to implanting a magnesium-based Magmaris bioresorbable scaffold (MgBRS) on lumen measurements using optical coherence tomography (OCT) and on clinical outcomes after 12 months. Method: In the OPTIMIS-study (Optimal lesion preparation before implantation of a MgBRS in patients with coronary artery stenosis), patients were randomly assigned to pre-dilatation with a scoring balloon or a standard non-compliant balloon before MgBRS implantation. OCT was performed before and after scaffold implantation, and at 6- and 12-month follow-up. Clinical 12-month follow-up was performed for all 82 enrolled patients. Results: Serial pre- and post-procedural, 6- and 12-month OCT were available in 33 lesions (scoring: n = 16 vs. standard: n = 17). Minimal lumen area (MLA) (scoring 4.6 ± 0.5 mm2 vs. standard 3.6 ± 0.5 mm2, p = 0.16) did not differ significantly at 12 months. In both the scoring and standard non-compliant balloon group, MLA changed significantly over time (post-procedure – 6-month – 12-month) (scoring balloon: 7.2 ± 1.8 mm2–5.1 ± 1.7 mm2–5.0 ± 1.9 mm2, p < 0.01, and standard non-compliant balloon: 6.2 ± 1.1 mm2–5.2 ± 1.5 mm2–5.3 ± 2.0 mm2, p < 0.01). No target lesion revascularizations (TLR) occurred in the scoring balloon group, compared to 7 TLR in the standard non-compliant balloon group within 12 months. Conclusion: In lesions treated with MgBRS, MLA did not differ significantly after 12 months among lesions pre-dilated with a scoring balloon compared to a standard non-compliant balloon. However, only lesions pre-dilated with a scoring balloon were free from TLR.
AB - Background: Lumen reduction after bioresorbable scaffold implantation has been reported. This study aimed to assess the influence of pre-dilatation with a scoring balloon versus a standard non-compliant balloon prior to implanting a magnesium-based Magmaris bioresorbable scaffold (MgBRS) on lumen measurements using optical coherence tomography (OCT) and on clinical outcomes after 12 months. Method: In the OPTIMIS-study (Optimal lesion preparation before implantation of a MgBRS in patients with coronary artery stenosis), patients were randomly assigned to pre-dilatation with a scoring balloon or a standard non-compliant balloon before MgBRS implantation. OCT was performed before and after scaffold implantation, and at 6- and 12-month follow-up. Clinical 12-month follow-up was performed for all 82 enrolled patients. Results: Serial pre- and post-procedural, 6- and 12-month OCT were available in 33 lesions (scoring: n = 16 vs. standard: n = 17). Minimal lumen area (MLA) (scoring 4.6 ± 0.5 mm2 vs. standard 3.6 ± 0.5 mm2, p = 0.16) did not differ significantly at 12 months. In both the scoring and standard non-compliant balloon group, MLA changed significantly over time (post-procedure – 6-month – 12-month) (scoring balloon: 7.2 ± 1.8 mm2–5.1 ± 1.7 mm2–5.0 ± 1.9 mm2, p < 0.01, and standard non-compliant balloon: 6.2 ± 1.1 mm2–5.2 ± 1.5 mm2–5.3 ± 2.0 mm2, p < 0.01). No target lesion revascularizations (TLR) occurred in the scoring balloon group, compared to 7 TLR in the standard non-compliant balloon group within 12 months. Conclusion: In lesions treated with MgBRS, MLA did not differ significantly after 12 months among lesions pre-dilated with a scoring balloon compared to a standard non-compliant balloon. However, only lesions pre-dilated with a scoring balloon were free from TLR.
KW - Bioresorbable scaffold
KW - Optical coherence tomography
KW - Stable angina
U2 - 10.1016/j.carrev.2025.03.003
DO - 10.1016/j.carrev.2025.03.003
M3 - Journal article
C2 - 40074646
AN - SCOPUS:105000080585
SN - 1553-8389
VL - 77
SP - 45
EP - 53
JO - Cardiovascular Revascularization Medicine