TY - JOUR
T1 - The impact of inter-laboratory glucose bias on the diagnosis of gestational diabetes mellitus
T2 - Comparison of common automated central laboratory methods
AU - Scheuer, Cathrine Munk
AU - Tvarnø, Casper Duevang
AU - Gils, Charlotte
AU - Ravn, Julie Dahl
AU - David McIntyre, H
AU - Jensen, Dorte Møller
AU - Damm, Peter
AU - Lauenborg, Jeannet Kepp Bruun
AU - Clausen, Tine Dalsgaard
AU - Overgaard, Martin
N1 - Copyright © 2023 The Authors. Published by Elsevier B.V. All rights reserved.
PY - 2023/6
Y1 - 2023/6
N2 - BACKGROUND AND AIMS: The diagnosis of gestational diabetes mellitus (GDM) is based exclusively on glucose measurements, which are highly influenced by pre-analytical and analytical factors. Therefore, poor agreement across laboratories may affect the prevalence of GDM. We aimed to determine the inter-laboratory bias of glucose measurements and the impact on GDM prevalence.MATERIAL AND METHODS: A prospective cohort study of women (n = 110) referred for second-trimester GDM diagnostics using a 75 g oral glucose tolerance test. Maternal glucose was assessed from venous plasma at fasting, 1 h and 2 h. Venous blood were collected in Fluoride Citrate tubes and frozen. Samples were analyzed at five central laboratories using four different automated glucose Hexokinase methods and GDM prevalence was evaluated according to WHO2013 diagnostic criteria.RESULTS: Maximum inter-laboratory bias was 0.19, 0.30 and 0.27 mmol/L in fasting, 1 h and 2 h samples, respectively. GDM prevalence ranged 30.0-41.1% across laboratories.CONCLUSION: Inter-laboratory bias for mean venous glucose was low and within desirable limits. Nonetheless, the impact on GDM prevalence was considerable, which may inappropriately affect clinical practice.
AB - BACKGROUND AND AIMS: The diagnosis of gestational diabetes mellitus (GDM) is based exclusively on glucose measurements, which are highly influenced by pre-analytical and analytical factors. Therefore, poor agreement across laboratories may affect the prevalence of GDM. We aimed to determine the inter-laboratory bias of glucose measurements and the impact on GDM prevalence.MATERIAL AND METHODS: A prospective cohort study of women (n = 110) referred for second-trimester GDM diagnostics using a 75 g oral glucose tolerance test. Maternal glucose was assessed from venous plasma at fasting, 1 h and 2 h. Venous blood were collected in Fluoride Citrate tubes and frozen. Samples were analyzed at five central laboratories using four different automated glucose Hexokinase methods and GDM prevalence was evaluated according to WHO2013 diagnostic criteria.RESULTS: Maximum inter-laboratory bias was 0.19, 0.30 and 0.27 mmol/L in fasting, 1 h and 2 h samples, respectively. GDM prevalence ranged 30.0-41.1% across laboratories.CONCLUSION: Inter-laboratory bias for mean venous glucose was low and within desirable limits. Nonetheless, the impact on GDM prevalence was considerable, which may inappropriately affect clinical practice.
U2 - 10.1016/j.cca.2023.117414
DO - 10.1016/j.cca.2023.117414
M3 - Journal article
C2 - 37263535
SN - 0009-8981
VL - 546
JO - Clinica Chimica Acta
JF - Clinica Chimica Acta
M1 - 117414
ER -