Falsely negative urinary leucocyte counts due to delayed examination

H. Kierkegaard, U. Feldt-Rasmussen, M. Hørder*, H. J. Andersen, P. J. Jørgensen

*Kontaktforfatter

    Publikation: Bidrag til tidsskriftTidsskriftartikelForskningpeer review

    Abstract

    Microscopic examination for urinary leucocytes was performed in the ward immediately after micturition and approximately 3 h later at a laboratory. Of the forty-eight specimens classified as positive by counting in the ward, 35% erroneously turned up negative by later counting at the laboratory. Counting of urinary leucocytes should consequently be performed as soon as possible after micturition, probably within 1 h.

    OriginalsprogEngelsk
    TidsskriftScandinavian Journal of Clinical and Laboratory Investigation
    Vol/bind40
    Udgave nummer3
    Sider (fra-til)259-261
    ISSN0036-5513
    DOI
    StatusUdgivet - 1. jan. 1980

    Fingeraftryk

    Dyk ned i forskningsemnerne om 'Falsely negative urinary leucocyte counts due to delayed examination'. Sammen danner de et unikt fingeraftryk.

    Citationsformater